E. coli and most bacteria
Lysates for purifying recombinant proteins, enzymes and nucleic acids. E. coli cells break at moderate pressures.
- E. coli
- Recombinant proteins
- Enzymes
The C-series homogenizers are designed to disrupt cells: bacteria, yeasts, spores. Here is what they do, their limits, and what we check with you before proposing a model.
The sample, in suspension, is pressurized and then forced through the narrow gap of the valve. The sudden pressure drop, shear and impact against the valve's impact ring break the cell walls. Repeating the pass increases the disruption rate.
Lysates for purifying recombinant proteins, enzymes and nucleic acids. E. coli cells break at moderate pressures.
Yeasts have a thicker wall: they need higher pressure and several passes. The manufacturer reports up to three passes to exceed 99 % disruption.
The manufacturer lists the disruption of Ganoderma spores among the uses of the C series, alongside yeasts and E. coli.
The manufacturer reports a disruption rate of 95 to 99 % on yeasts, E. coli and Ganoderma spores. The rate depends on the organism, concentration, pressure and number of passes.
The manufacturer's manual calls for at least two passes. For yeast, its tests show up to three passes to exceed 99 %. For E. coli, its datasheets show tests with one or two passes between 800 and 1,800 bar.
During the demonstration, we process your sample at several pressures and numbers of passes. You measure the result with your usual method: microscope counting, optical density or assay of released proteins.
These limits come from the manufacturer's manual. We check them with you before any demonstration.
Dry matter must not exceed 20 % of the suspension. Above that, the sample must be diluted.
Describe your sample: we will tell you whether the C series fits, and with which model.